EFFICIENT ABLATION OF EX VIVO CORNEA USING THE FIFTH HARMONIC OF FEMTOSECOND YB:KGW LASER
E. Gabrytė1, 2, M. Vengris1, R. Danielius2, O. Rukšėnas3, A. Vaičeliūnaitė3
1 Laser Research Centre, Faculty of Physics, Vilnius University, Lithuania
2 Light Conversion Ltd, Lithuania
3 Faculty of Natural Sciences, Vilnius University, Lithuania
Corrective eye surgery is a rapidly developing and promising field of ophthalmology [1, 2]. Processes of corneal flap creation and shaping of cornea are currently performed using two different lasers (ArF excimer and femtosecond solid-state), or excimer laser in combination with microkeratome [3]. In this study, we aim to develop an universal solid-state femtosecond laser that can be used during both stages of the surgery.
Here we presented the system of all-solid-state femtosecond laser and a fifth harmonic generator and its applications to the ablation of transparent materials, including PMMA plastic, gelatine and ex vivo porcine corneas (Fig. 1.).
Fig. 1. Results of ex vivo porcine corneas shaping. (A) Surface topographic image after ablation; (B) cross-section of corneal surface before and after shaping, resulting a positive 10 D refraction change.
Spherical structures were produced in samples of these materials. Resulting surfaces possessed a fine optical quality. Corneal shaping to the 12 D refraction change lasts ca. 180 s using current system. The presented system has a great potential to be applied in different areas of ophthalmic laser surgery, because it produces high-intensity femtosecond laser pulses at 1027, 514 and 205 nm.
[1] I. G. Pallikaris et al., Lasers in Surgery and Medicine, **10(5)**: 463-468 (1990).
[2] S.L. Trokel, R. Srinivasan, and B.B. Braren, American Journal of Ophthalmology, **96**: 710-714 (1983).
[3] R. Montes-Mico, A. Rodriguez-Galietero, and J.L. Alio, Ophthalmology, **114(1)**, 62-68 (2007).