In high energy particle experiments it is very important to register ionising radiation rapidly, reliably and effectively. Therefore, Si pin particle detectors, operating at full depletion regime are commonly employed. However, radiation induced defects affect the detectors functional parameters by the increase of leakage current, of full depletion voltage, and by a reduction of charge collection efficiency, etc [1]. So it is important to investigate the variations of these parameters in radiation damaged diodes. Commonly, these parameters are examined by employing current-voltage (I-V) and capacitance-voltage (C-V) techniques [2].
In this work Float Zone (FZ) and Magnetic Czochralski (MCZ) Si pin diodes irradiated by reactor neutrons at fluences in the range of 1012-1016 cm-2 have been examined by combined analysis of I-V and C-V characteristics measured at various temperatures. Measurements of I-V and C-V characteristics have been performed by employing Sub-Femtoamp Keithley 6430 electrometer and high precision QuadTech 7600 LRC meter, respectively. Temperature dependent measurements have been carried out by placing the samples in the cryostat.
It has been found that leakage current increases proportionally to a fluence increment of reactor neutrons irradiation at room temperature in FZ and MCZ Si diodes due to the radiation induced defects acting as generation centres. The capacitance obtained at low temperatures (<120 K) in irradiated detectors is invariable with voltage and equal to a value of a diode geometrical capacitance. This indicates a complete compensation of the shallow dopants by radiation traps within a range of low temperatures. It was also shown that different kind defects are produced in FZ and MCZ Si diodes during irradiation by close values of neutron fluence.