PHOTO-IONIZATION SPECTROSCOPY OF DEEP LEVELS IN Si RADIATION DETECTORS

Aušra Čerkauskaitė1, Vidmantas Kalendra1, 2, Eugenijus Gaubas2

1 Faculty of Physics, Vilnius University, Saulėtekio 9-III, LT-10222 Vilnius, Lithuania

2 Institute of Applied Research, Vilnius University, Saulėtekio 9-III, Vilnius, Lithuania

Silicon is the most widely used material for fabrication of particle detectors [1]. The main problem arises due to low radiation tolerance of the Si detectors under irradiation fluences of hadrons above 1015 particle/cm2. Thus, an expensive procedure of the detectors replacement is inevitable. The main objective of this work is to analyse the deep level spectra in the as-fabricated in Si detectors as well as irradiated with high fluencies of hadrons and to study variations of the deep centres densities.

The photo-ionization spectroscopy instruments based on recording of the microwave probed pulsed photoconductivity and of the steady-state current within diode structures have been installed and approved. The diode structure sample is placed into vacuumed cryostat and cooled by liquid nitrogen, to record photo-ionisation spectra with reduced electrical noises. Photo-ionisation is implemented by a halogen lamp biased by dc voltage. The DMR-4 monochromator is exploited to disperse the excitation light. This monchromator is placed in between the cryostat and the halogen lamp. Changing of excitation wavelength is handled by a step-motor which is connected through Line Print Terminal (LPT) to a parallel port interface of a personal computer. A platinum thermo-resisting sensor is used for the temperature measurements on the sample. This sensor signal is measured by the Agilent 34401A multi-meter. Current measurements are performed by employing Keithley 6487 pico-amper-meter with internal DC source. The data sampling and processing is performed by personal computer using Labview System Design Software.

The spectral scans were performed by increasing excitation photon energy to avoid undesirable emptying of the shallower levels before photo-ionisation during record of spectrum.

The photo-ionization spectra were analyzed by employing the Lucovsky model [2]. According to this model the photocurrent (I) is given by:

$$I \sim n_M \Delta E_M^{0.5} (h \nu - \Delta E_M)^{1.5} / (h \nu)^3$$

where $\Delta E_M$ is the optical activation energy of the deep centre, $h\nu$ is the photon energy, $n_M$ is the initial trap filling.

The optical activation energy values obtained from the photo-ionization spectra can be related with the excitation of electrons from the filled deep centres or generation of free holes by excitation of electrons from the valence band into the empty deep levels.


[1] www.cern.ch

[2] G. Lucovsky, Solid State Comm, 3 (.1965) 299.