A SYSTEM FOR MEASURING SURFACE ROUGHNESS BY TOTAL INTEGRATED SCATTERING

Lina Mažulė1, Simona Liukaitytė1, Valdas Sirutkaitis1

1 Laser Research Centre, Vilnius University, Sauletekio 10, 10223 Vilnius, Lithuania

Surface roughness, contaminations, micro scratches, fractures, various defects and fluctuations of the density in the bulk of optical component reduce quality of the components and increase scattering of the incident light [1]. There are several techniques for testing surface topography. The Nomarski microscope, atomic-force microscope, electron microscope, scanning tunneling microscope and other microscopes are used for direct investigation.

Light scattering measurements are indirect methods for testing the topography of a surface. These measurements allow the rapid evaluation of larger surface areas [2, 1]. The theory of light scattering enables us to calculate the surface roughness from light scattering, and the measurements can have resolution appropriate for the evaluation of super-polished surfaces (the roughness of surfaces in the range of few Å).

A system for measurement of surface roughness based on total integrated scattering (TIS) at 1064nm, 532 nm and 355 nm is built and demonstrated (Fig. 1). Surfaces up to 25 mm×25 mm are scanned in 6 minutes with a spatial resolution of 0.4 mm. Careful attention to reducing stray light and purging the measurement chamber with filtered air allow scattering resolution better than 10-5 (Table 1). Surface roughness measurements better than 1 nm RMS are demonstrated and confirmed by comparison measurements with an atomic-force microscope [3].

Figure 1
Fig. 1. Total integrated scattering measurement setup (1 - laser with harmonic generators, 2 – the other laser (wavelengths: 1030, 515 and 343 nm), 3 – attenuator, 4, 5, 9 and 11 mirrors, 6, 7 and 8 – filters, 10 - 1 m focal lens, 12 – aperture, 13 – Ulbricht sphere, 14 - translation stage, 15 - detector, 16 – beam dumps, 17- the chamber of measurements, 18 - clean air flow, 19 - return air flow, 20 – computer).
Table 1. The scattering measurement resolution for different wavelengths
Wavelength, nm:1064532355
The scattering signal without the sample, ppm:4.2±0.14.5±0.59.6±0.9
Figure 2
Fig. 2. The surface roughness of high reflectance mirror at 1064 nm AOI=0 deg.

[1] Amra C, Torricini D and Roche P 1993 Multiwavelength (0.45-10.6 μm) angle-resolved scatterometer or how to extend the optical window Appl. Opt. 32 5462-74

[2] Bennett J M and Mattsson L 1999 Introduction to surface roughness and scattering (Washington, D.C.: Optical Society of America)

[3] L Mazule S L, R C Eckardt, A Melninkaitis, O Balachninaite, V Sirutkaitis 2011 A system for measuring surface roughness by total integrated scattering Journal of Physics D: Applied Physics 44 9