ANALYSIS OF CELL ELECTROPORATION INDUCED BY ELECTRIC PULSE WITH THE DURATION FROM 40 NS TO 2 MS

Aiste Bitinaite1, 2, Rita Saule1, Voitech Stankevic2, Nerija Zurauskiene2, Saulius Balevicius2, Gintautas Saulis1, 2

1 Department of Biology, Vytautas Magnus University, Lithuania

2 Laboratory of High Power Pulses, Institute of Semiconductor Physics, Lithuania

To optimize practical applications of electroporation, it is important to know whether and how many of cells have become electroporated under the influence of a particular electric field treatment. However, to determine whether the cell is porated or still intact is rather a complicated task. The matter is that different treatments create different pore populations: the size of the pores and their lifetime depend on the duration of the pulse.

Here, the size and lifetime of the pores created by a single electric field pulse of different durations have been evaluated. The fraction of electroporated mouse hepatoma MH-22A and Chinese hamster ovary cells was determined from the extent of the release of intracellular potassium ions and erythrocytes - from the extent of their hemolysis after 20-24 h-incubation in 0.63% NaCl solution at 4°C. The size of the pores was evaluated by studying membrane permeability to bleomycin and protective action of xylitol, mannitol, and sucrose against colloid-osmotic lysis.

In the case of a short pulse (square-wave with the duration 40 ns-100 μs), in the large portion (30–60%) of electroporated (permeable to potassium ions) cells, an electric pulse created only the pores, which were smaller than the molecules of bleomycin (radius $\approx$ 0.8 nm) or sucrose (radius 0.44–0.52 nm). In the case of a long pulse (duration 1-2 ms), in almost all cells, which were electroporated, there were the pores larger than the molecules of bleomycin and/or sucrose. Kinetics of pore resealing depended on the pulse duration as well and was faster after shorter pulses.

The results obtained here show that the methods based on measuring membrane permeability for rather large tracer molecules, e.g. propidium iodide or trypan blue, which are suitable for determination of cell electroporation by long pulses, can fail in the case of short pulses.