Biosensors are analytical devices used mainly for the recognition of the chemical substances in solution to be analyzed [1]. The biosensors usually utilize specific biochemical reactions catalyzed by enzymes immobilized on electrodes. Allosteric enzymes are unique compared to other enzymes because of its multiple interacting active sites. These enzymes are very attractive in different applications, e. g. for detecting disease [2]. Mathematical modelling is one of the most attractive solutions to understanding the peculiarities of the biosensors and optimizing their configuration [3].
The purpose of this work was to develop a mathematical as well as the corresponding computational models of a biosensor based on an allosteric enzyme and to investigate the influence of the allosteric effect on the biosensor response. The developed mathematical model of the biosensor is based on non-stationary diffusion equations and involves a model of the allosteric reaction network [4]. The model comprises two compartments: a layer of the enzyme entrapped on the electrode surface and an outer diffusion layer modelled by the Nernst approach [1, 3]. The governing equations in the dimensionless form read as follows:
$$ \frac{\partial S_e}{\partial T} = \frac{\partial^2 S_e}{\partial X^2} - \sigma^2 \frac{S_e(K_m' + kS_e)}{K_m' + K_m' S_e + S_e^2} \ \frac{\partial P_e}{\partial T} = D_{P_E} \frac{\partial^2 P_e}{\partial X^2} + \sigma^2 \frac{S_e(K_m' + kS_e)}{K_m' + K_m' S_e + S_e^2} \tag{1} $$where $X$ is space, $T$ – time, $S_e(X,T)$ and $P_e(X,T)$ are the concentrations of the substrate and product, respectively, $K'_m$ is normalized Michaelis constant, $k$ is the allosteric ratio, and $\sigma^2$ is the diffusion module. The governing equations (1) together with appropriate initial, boundary and matching conditions form an initial boundary value problem, which was numerically solved by applying the finite difference technique [3].
By varying the input parameters, the output results were numerically analyzed with a special emphasis to the influence of the model parameters on the allosteric effect of the biosensor response. Numerical experiments showed that the allosteric effect nonlinearly increases with decreasing the enzyme concentration as well as decreasing the thickness of the enzyme layer.