INFLUENCE OF MOBILE PHASE COMPOSITION ON RETENTION AND EFFICIENCY IN HYDROPHILIC INTERACTION CHROMATOGRAPHY

Inga Baškirova1, Vilma Olšauskaitė1, Audrius Padarauskas1

1 Department of Analytical and Environmental Chemistry, Vilnius University, Lithuania

Hydrophilic interaction chromatography (HILIC) is a feasible alternative for the analysis of highly polar and ionized compounds that are poorly or even not retained in reversed-phase chromatography [1]. This separation technique uses a polar stationary phase (for example, unmodified silica or a polar bonded phase) in conjunction with a polar mobile phase containing more than 60-70% organic solvent (typically acetonitrile) in aqueous buffer. Although HILIC has been widely applied, the retention mechanism in this separation mode is still under debate [2]. The primary retention mechanism is believed to be partitioning of the analytes between a water layer adsorbed on the surface of the hydrophilic stationary phase and the less polar bulk mobile phase. However, secondary interactions such as hydrogen bonding, ionic and even hydrophobic can also occur depending on the nature of the analyte, stationary and mobile phases [3]. If the sample contains a large number of the analytes having different properties then it becomes difficult to predict their chromatographic behaviour.

The aim of the current work was to investigate the effect of mobile phase composition on the chromatographic behaviour of anionic, cationic and neutral compounds under hydrophilic interaction chromatography conditions using bare silica stationary phase.

HILIC separations were performed on a Waters Acquity UPLC system (Waters, Milford MA, USA) equipped with an Acquity UPLC photodiode array detector (PDA). Acquity UPLC BEH HILIC column (2.1×100 mm, 1.7 μm, Waters) was used in the experiments. We selected six compounds (acetylsalicylic acid, creatinine, nicotine, nicotinic acid, acetaminophen and uracil) as model analytes to examine the retention and efficiency properties of the mobile phases.

The effect of three organic solvents, namely acetonitrile (ACN), methanol (MeOH) and isopropanol (IPA) on the chromatographic behaviour of the analytes was investigated. For all analytes the eluting strength follows the order: ACN < IPA < MeOH, which is correlated with the different hydrogen bonding abilities of the solvents. Polar protic solvents (MeOH and IPA) can be both donor and acceptor of hydrogen bonds and aprotic ACN can be only hydrogen bond acceptor. Consequently, protic solvents can more effectively compete for polar active sites on the surface of stationary phase, replacing water molecules and thus producing a more hydrophobic surface. For all analytes acetonitrile provided the highest peak efficiency.

As expected, all coumpounds exhibited typical HILIC behaviour of increasing retention with increasing ACN content in the mobile phase. The retention for the charged analytes were much more strongly affected by ACN concentration than for the neutral compounds. This is possibly due to enhanced secondary analyte/stationary phase interactions (e.g., hydrogen bonding and electrostatic interactions) which may become predominant as water content decreases.

Minimal changes in retention and efficiency of neutral compounds were observed upon changing pH (3-7) and ammonium acetate buffer concentration (0-10 mmol/L). The efficiency of charged analytes significantly increased with increasing pH and buffer concentration. The changes in retention for cationic and anionic compounds with increasing pH and buffer concentration were in the opposite direction. The retention of anions increased, whereas an opposite trend was observed for cationic analytes.

Obtained results showed that the three mobile phase parameters (concentration of organic solvent, pH and buffer concentration) enabled manipulation of the retention and efficiency of analytes with different charge state and should be useful for optimization of mobile phase during HILIC method development.


[1] B. Buszewski, S. Noga, Hydrophilic interaction liquid chromatography (HILIC) - a powerful separation technique, Anal. Bioanal. Chem. 402, 231-247 (2012).

[2] P. Hemstrom, K. Irgum, Hydrophilic interaction chromatography, J. Sep. Sci. 29, 1784-1821 (2006).

[3] A. E. Karatapanis, Y. C. Fiamegos, C. D. Stalikas, A revisit to the retention mechanism of hydrophilic interaction liquid chromatography using model organic compounds. J. Chromatogr. A 1218, 2871-2879 (2011).