THE INFLUENCE OF RADON CONCENTRATION ON THE BACKGROUND SPECTRUM MEASURED BY GAMMA SPECTROSCOPY.

Damian Kasyan, Krzysztof Kozak

1 Department of Physics, Cracow University of Technology, Poland

2 Laboratory of Radiometric Expertise, Institute of Nuclear Physics Polish Academy of Sciences, Poland

The background is very important problem in gamma spectroscopy. The accurate determination of the background guarantees more accurate result of the measurement. The reduction of the background makes it possible to lower minimum detectable activity. Unfortunately, the environmental radioactivity normally present in laboratory influences the level of background.

The main part of this radioactivity comes from building materials and radon progeny. Radon's chemical and physical properties cause its migration into laboratory building, especially located in lower floors.

The aim of this work was to study how radon concentration may influence the background gamma spectrum. Using a spectrometer with HPGe detector the background gamma spectra were collected in different radon concentrations which were continuously registered by a professional radon monitor AlphaGUARD PQ2000 PRO. Radon concentration in laboratory was changed with the application of certified radon source PYLON 1025. The comparison of the spectra is presented as well as the method for decreasing radon concentration inside the HPGe detector's shielding. This method is based on the ventilation with liquid nitrogen vapors.

The work was performed at the Laboratory of Radiometric Expertise, Institute of Nuclear Physics PAN, Krakow, Poland.