Open Readings 2026 • P2-82
DEVELOPMENT OF A MULTILAYERED, MOLECULARLY IMPRINTED POLYMER-BASED ELECTROCHEMICAL SENSOR FOR TETRACYCLINE DETECTION
Beatričė Kulikauskaitė1, Domas Pirštelis1, 2, Ernestas Brazys1, 2, Vilma Ratautaitė2, Arūnas Ramanavičius1, 2
1 Vilnius University, Faculty of Chemistry and Geosciences, Institute of Chemistry, Department of Physical Chemistry, Lithuania
2 State Research Institute Center for Physical Sciences and Technology, Department of Nanotechnology, Lithuania
The widespread use of antibiotics, such as tetracycline, has resulted in persistent drug residues in the environment, which require monitoring to evaluate pollution levels. Molecularly imprinted polymer (MIP)–based sensors are considered a robust, low-cost, and easy-to-use solution for the development of analytical systems for the detection of various analytes [1,2]. The introduction of interfacial underlayers, such as self-assembled monolayers, between the electrode and the MIP layer enables the tuning of MIP properties [3].
In this work, a multilayered MIP sensor was constructed on a screen-printed carbon electrode (SPCE) platform. The SPCE was modified by drop-casting a poly(3,4-ethylenedioxythiophene):polystyrene sulfonate (PEDOT:PSS) mixture as an underlayer to tune charge-transfer characteristics. Subsequently, a molecularly imprinted polypyrrole (MIP-Ppy) layer, using tetracycline as the template, was electrochemically synthesized on the PEDOT:PSS/SPCE surface. A non-imprinted polypyrrole (NIP-Ppy) layer was prepared under identical conditions to demonstrate imprinting selectivity. The rebinding of tetracycline to the MIP/PEDOT:PSS/SPCE sensor was carried out in an acetate medium and was evaluated using electrochemical impedance spectroscopy.
The constructed sensor enabled the detection of tetracycline at micromolar concentrations. The proposed multilayered, easy-to-fabricate, MIP-based electrochemical sensor demonstrates promising potential for application in the environmental monitoring of antibiotic residues.
Acknowledgments: This project has received funding from the Research Council of Lithuania (LMTLT), agreement No. S-LL-25-3.
[1] E. Brazys, V. Ratautaite, E. Mohsenzadeh, R. Boguzaite, A. Ramanaviciute, and A. Ramanavicius, “Formation of molecularly imprinted polymers: Strategies applied for the removal of protein template (review),” Advances in Colloid and Interface Science, vol. 337, p. 103386, Dec. 2024, doi: 10.1016/j.cis.2024.103386.
[2] S. Ramanavicius, U. Samukaite-Bubniene, V. Ratautaite, M. Bechelany, and A. Ramanavicius, “Electrochemical molecularly imprinted polymer based sensors for pharmaceutical and biomedical applications (review),” Journal of Pharmaceutical and Biomedical Analysis, vol. 215, p. 114739, Mar. 2022, doi: 10.1016/j.jpba.2022.114739.
[3] V. Liustrovaite, V. Ratautaite, A. Ramanaviciene, and A. Ramanavicius, “Detection of the SARS-CoV-2 nucleoprotein by electrochemical biosensor based on molecularly imprinted polypyrrole formed on self-assembled monolayer,” Biosensors and Bioelectronics, vol. 272, p. 117092, Dec. 2024, doi: 10.1016/j.bios.2024.117092.