Methods that allow the identification of protein biomarkers can potentially make a significant contribution to biomedical applications and individual health. In our work, several analytical methodologies using carbon based electrodes modified with various nanostructures were reported [1-4]. However, the sensing of biomarkers related to Coronavirus disease (COVID-19) is still in demand; the World Health Organization in 2020 has declared COVID-19 as a global pandemic. Because of the presence of asymptomatic carriers, nucleic acid testing for diagnosing of COVID-19, may result in false negatives and is difficult to apply for every suspected patient. Meanwhile, detecting specific antibodies in blood against the Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) virus, such as the Immunoglobulin M (IgM) antibody, is another option for COVID-19 diagnosis (including different strains), as IgM is widely accepted as an important indicator during the acute infection period, and can be detected within 5-7 days since infection.

Through the indirect electrochemistry method, a colloidal gold nanoparticle-based novel assay was developed to achieve rapid diagnosis and on-site detection of the IgM antibody against the SARS-CoV-2 virus.

The aqueous synthesis of gold nanoparticles (AuNPs) was carried out using 'green and sustainable' reducing agents. The effect of various variables such as temperature and the presence of nontoxic capping agents on the reduction of auric chloride (Au (III)) salt and the growth of AuNPs was investigated using UltraViolet-Visible Spectroscopy (UV-VIS), Raman spectroscopy (RS), transmission electron microscopy (TEM).

Subsequently, the SARS-CoV-2 spike protein was conjugated with AuNPs for IgM capturing to form an electrochemical transducer for electrochemical sensing of IgM. Physiological concentrations of IgM were detected using screen printed carbon electrodes. For IgM detection, different electrochemistry methods were used, including differential pulse voltammetry (DPV), electrochemical impedance spectroscopy (EIS), and fast Fourier transform electrochemical impedance spectroscopy (FTT-EIS).

Additionally, blood serum samples were used to assess the performance of the AuNP assay, and the experimental results will be presented.

Acknowledgments: Dr Ausra Baradoke is grateful to the European Social Fund (No. 09.3.3-LMT-K-712-23-0159) under a grant agreement with the Research Council of Lithuania (LMTLT). Alexandra Elsakova and Ali Jafarov are thankful to the Erasmus+ funding program.


[1] A. Baradoke, R. Hein, X. Li, J.J. Davis. Reagentless Redox Capacitive Assaying of C-Reactive Protein at a Polyaniline Interface. Anal. Chem. 92, 3508-351, 2020.

[2] A. Baradoke, I. Pastoriza-Santos, E. González-Romero. Screen-printed GPH electrode modified with Ru nanoplates and PoPD polymer film for NADH sensing: Design and characterization. Electrochim. Acta 300, 2019.

[3] A. Baradoke, B. Jose, R. Pauliukaite, R. J. Forster. Properties of Anti-CA125 antibody layers on screen-printed carbon electrodes modified by gold and platinum nanostructures. Electrochim. Acta 306, 299–306, 2019.

[4] A. Baradoke, A. Santos, P. R. Bueno, J.J. Davis Introducing polymer conductance in diagnostically relevant transduction. Biosens. Bioelectron. 112705, 2020.