Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is a virus that causes COVID-19 disease. The COVID-19 pandemic began in late 2019 and is still ongoing. SARS-CoV-2 virus contains four structural proteins: spike, nucleocapsid, envelope and membrane [1]. In this study SARS-CoV-2 spike glycoprotein was chosen as a target for the developed sensor system. Described research is dedicated to the design of electrochemical SARS-CoV-2 virus protein detection systems based on the application of molecular imprinting technique.
The typical molecular imprinting procedure consists of chemical or electrochemical polymerization from a mixture containing functional monomers and template molecules. Following template removal by washing in a selected solvent, the polymer binding sites are left in the structure of polymer, which are specific or complementary to the template molecules [2]. This characteristic can be applied in the design of both biological and chemical molecule sensor systems [3].
In this research, SARS-CoV-2 spike glycoprotein imprinted polypyrrole-based (MIP-Ppy) sensor was developed [4]. The polymeric mixture consisted of 0.5 M pyrrole as a monomer and 50 μg/mL of SARS-CoV-2 spike glycoprotein as a template in PBS solution, pH 7.4. The polymeric layer was formed electrochemically onto the surface of a platinum electrode by a sequence of potential pulses. After the polymerization template molecules were extracted by washing the polymeric film in 0.05 M H2SO4 for 10 min. The characteristics of the electrode modified by the imprinted polypyrrole films were evaluated by pulsed amperometric detection. In comparison an electrode was modified with non-imprinted polypyrrole film (NIP-Ppy).
Obtained results indicate that the sensor based on MIP-Ppy has a higher sensitivity toward SARS-CoV-2 glycoprotein than the electrode modified by NIP-Ppy film.