DEVELOPMENT AND TESTING OF CRYOFOCUSING ADD-ON FOR GAS CHROMATOGRAPHY

Audrius Sadaunykas1, Audrius Zolumskis1, Evaldas Naujalis1

1 Metrology Department, State Research Institute Center for Physical Sciences and Technology Lukiškių str. 9, LT-01108 Vilnius, Lithuania

Gas chromatography (GC) is one of the most popular analytical method used around the world. This method is applied for various volatile compound analysis. Basics of gas chromatography is compound separation, performed on a chromatographic GC column. Analytes are pushed forward along the column (a thin silica capillary tube, covered in stationary phase) using inert gas flow and because of chemical forces and bonds between analytes and stationary phase of the column, analytes are retained and elute from the column at different rates [1]. GC requires a very precise temperature control to influence analyte separation and make analysis more efficient. Lack of sensitivity and not enough separation are the main problems researchers face while using gas chromatography. There are different ways to improve sensitivity and separation: changes in detector, column or injection system [2]. In our laboratory we are building an add-on, which could be used with existing GC systems, without any modification to existing parts.

Our prototype scheme showed in Figure 1 is based on cooling down part of the column to -60 °C. Unit outside the GC houses cylindrical aluminum block, which is cooled down using external chiller. On top of the block there is a smaller disk-shaped part, that houses the GC column and resistive heaters. This part can be lifted up from the block heated to 100 °C and then, lowered down to cool down quickly. By having part of the column at -60 °C allows us to trap the analyte traveling from the injector and contain it there. Multiple injections could be made and analyte peak area enriched. After number of injections, disk is raised and heater is turned on. Temperature in the disk reaches 100 °C and analyte is eluted quickly. Number of injections corelate with increase in peak area of the analyte. This process greatly increases the sensitivity of the system.

To test the accuracy and reproducibility of our add-on paired with GC system, we decided to measure nonane (1 μl) dissolved in pentane (1 ml) at different enrichment levels. This produced a curve (Figure 2) with R2 of 0,9999. This indicates, that all injections are consistent and no loss of analyte is present.

Figure 1 & 2
Fig. 1. Gas chromatography system with cryofocusing add-on
Fig. 2. Peak areas at different enrichment levels

These results are just the beginning of our work on this topic and we will continue with improvement of the prototype.


[1] Skoog, Douglas A.; West, Donald M.; James Holler, F.; Crouch, Stanley R., Fundamentals of analytical chemistry, (Ninth ed.). Belmont, CA, (2013) ISBN-13: 978-0495558286

[2] Schomburg G., Practical Limitations of Capillary Gas Chromatography, Journal of High Resolution Chromatography, 2(7), 461-474, (1979).

[3] Sadaunykas A., On-column cryofocusing and analyte enrichment device for gas chromatography systems, Chemija, Vol 31, Issue 4, 284-289, (2020)