Procedure of the X-ray diagnostics have an essential role in fields of epidemiology, traumatology and oncology. Doctors at mentioned areas often use it for the diagnosis and visual presentation of the disease picture. In order to minimize harmful affection on patients it's necessary to most accurately account contribution of the X-ray dose received during medical procedures to annual dose of radiation.
During radiological procedure on patient one can estimate whether a used radiation dose is critically high for a current procedure under normal conditions. A parameter called diagnostic reference level DRL is used for that estimation. Equivalent parameter to the DRL is the dose accumulation coefficient $D_{acc}$ which characterizes the ratio of energy obtained from the both reflected and direct radiation to energy of the direct one. The $D_{acc}$ is used in order to measure level of affection on the patient as a dose obtained at the patient effective surface during the X-ray research procedure.
This in silico research studies influence of the X-ray machine parameters and the absorbing body properties on the dose accumulation coefficient $D_{acc}$ in the case of the X-ray diagnostics of a rib thorax.
A model of the X-ray diagnostics includes the X-ray machine model, the absorbing body model and the detector plate model. Geometrically the model of the X-ray diagnostics consist of a point X-ray source (the X-ray machine model) that placed in front of a circular cylinder filled with water (the patient model), right behind the cylinder is placed an air plate (the detector plate model). The cylinder and the detector plate are placed on the axis of the beam from the source.
In this approach used a model of the static X-ray machine, which includes an X-ray source, an energy spectrum smoothing filter and a collimation system.
The patient model parameters like height and radius of the cylinder are depend on the age of the patient being examined. Thus, for each patient a refinement is made to the value of the coefficient $D_{acc}$ by approaching the source and changing the radiation field. The parameters of the cylinder are taken on the basis of the tables of mass and growth of patients given in the ICRP-89.
The radiation detector model has linear dimensions of 5x5 cm and a thickness of 1 mm. The material of the detector plate is air under normal conditions. The detector is placed as close as possible to the cylinder. So, the distance from the detector to the source depends just on the radius of the cylinder. In addition, another case of detector geometry is a set of detectors located on the beam axis, placed at equal intervals from the cylinder. That configuration allows to determine the rate of decline of the dose accumulation coefficient $D_{acc}$ with the distance.
List of parameters that was changing in the research:
Voltage of the X-ray tube (kV)
Voltage ripple (%)
Effective patient body thickness (mm)
Patient age (years)
Smoothing filter depth (mm)
In this approach data for X-ray energy spectra from the source was taking from TASMIP [2], and include dependents from aluminum filter, voltage and voltage ripple. Monte-Carlo modelling used for computation energy, that was delivered to detector by gamma-quanta. Results of the in silico experiments:
1. In this work, more than 500 values for the were calculated for various exposure parameters.
2. The contribution of the smoothing filter depth to the coefficient value is 9% when considering ripple at 10% and its absence. Thus, it should be born in mind that when using filters of various thicknesses, a noticeable increase in the dose accumulation coefficient occurs.
3. The contribution of the effective patient body thickness to the coefficient value is 0.8% when considering ripple at 10% and its absence.
4. The dose accumulation coefficient increases with increasing voltage. An increase in the coefficient values with an increase in thickness is 4–4.7% for various stresses.
5. The difference in the value of the coefficient for the ages of 15 and 18 years depending on gender varies from 0.5% to 3.5% depending on changes in various parameters of the radiographic procedure.
Results of this work can be used to assess the success of lung radiography. The obtained results allow to determine the input dose on the surface for most types of patients, and the effect on these values of such characteristics of the X-ray tube as the values of the anode voltage, the smoothing filter thickness and the voltage ripple.